GTMG3083P150VV-X.X-FW Open Frame/Internal Medical Power Supply

GTMG3083P150VV-X.X-FW
GTMG3083P150VV-X.X-FW, Medical Power Supply, Open Frame/Internal, Open Frame Medical Power Supply, , Input Rating: 90 ~ 264VAC, 127 ~ 370VDC (47 ~ 63Hz), 3 Pin connector
Mating Housing: JST VHR-3N or equivalent
Mating Contact: JST SVH-21T-P1.1 or equivalent, Output Rating: 150 Watts, Power rating with convection cooling (W) , 12-48V in 0.1V increments, Approvals: VCCI; Ukraine; RoHS; PSE; GOST-R; cULus; cUR; China RoHS; CE; WEEE; Double Insulation;GTMG3083P150VV-X.X-FW, Medical Power Supply, Open Frame/Internal, Open Frame Medical Power Supply, , Input Rating: 90 ~ 264VAC, 127 ~ 370VDC (47 ~ 63Hz), 3 Pin connector Mating Housing: JST VHR-3N or equivalent Mating Contact: JST SVH-21T-P1.1 or equivalent, Output Rating: 150 Watts, Power rating with convection cooling (W) , 12-48V in 0.1V increments, Approvals: VCCI; Ukraine; RoHS; PSE; GOST-R; cULus; cUR; China RoHS; CE; WEEE; Double Insulation;
Information
Model NumberGTMG3083P150VV-X.X-FW
Agency Documentshttp://www.globtek.info/certs/GTMG3083PWWWVV-X.X-FW/
CE EC-Declaration for MedicalContact GlobTek for more information
CE EC-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ec_declaration?id=a0Oa000000MijYvEAJ
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for MedicalContact GlobTek for more information
UKCA-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ukca_declaration?id=a0Oa000000MijYvEAJ
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://cs.globtek.info/Morocco_SDoC_Declaration?id=a0Oa000000MijYvEAJ
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0Oa000000MijYvEAJ
Prop 65 Statementhttps://www.globtek.com/r2/PH7CNf
REACH Declarationhttps://www.globtek.com/r2/eM2co8
Conflict Minerals Declarationhttps://www.globtek.com/r2/vZY8R5
Model Parameters
TypeOpen Frame/Internal
TechnologyOpen Frame Medical Power Supply
CategoryMedical Power Supply
Input Voltage Range (V)90 ~ 264VAC, 127 ~ 370VDC (47 ~ 63Hz)
Max Input Current (A)3.23A / 100VAC, 1.23A / 240VAC
Max Output Power Rating (W)150
Output Voltage Range (V)12-48
EfficiencyN/A
Ingress ProtectionIPX0
Size (mm)50.8x32.1x101.6 mm
EnclosureUser-added image

User-added image
Rating Table
Model Number (Final ordering #/Part No to be defined)Voltage (V)Current (A)Power (W)RFQ
GTMG3083P15012-FW*1212.5150
GTMG3083P15015-FW*1510150
GTMG3083P15024-FW*246.25150
GTMG3083P15048-FW*483.12149.76
* (Final ordering #/Part No to be defined)
Specifications
250W Open Frame Medical Power Supply

Features:
Universal AC input / Full range with active PFC
2"x 4" footprint
Medical safety approval: ANSI / AAMI ES 60601-1 / TUV EN 60601-1; IEC 60601-1 / 2xMOPP
100W convection cooling and 150W with 10.5 CFM forced air
Protection: Short circuit, Over load, Over voltage
High operation temperature up to 70°C Power OK signal
Built-in fan supply
Withstand 2G vibration test


ELECTRICAL SPECIFICATIONS:
Output
DC Voltage Range: 12V 15V 24V 48V
Rated Curren:t 12.5A 10A 6.25A 3.12A
Current Range (Convection): 0 ~ 8.3A 0 ~ 6.7A 0 ~ 4.2A 0 ~ 2.1A
Current Range (23.5 CFM Fan): 0 ~ 12.5 A 0 ~ 10A 0 ~ 6.2 5A 0 ~3.12A
Rated Power (Convection): 99.6W 100.5W 100.8W 100.8W
Rated Power (23.5 CFM Fan): 150W 150W 150W 150W
Ripple & Noise (Max.): 120mVp-p 150mVp-p 240mVp-p 480mVp-p
Voltage Adj. Range: ±10%
Voltage Tolerance: ±2.0% ±2.0% ±1.0% ±1.0%
Line Regulation: ±0.5% ±0.5% ±0.5% ±0.5%
Load Regulation: ±2.0% ±2.0% ±2.0% ±2.0%
Setup, Rise Time: 2000ms, 50ms / 230VAC, 3500ms, 50ms / 115VAC at full load
Hold Up Time: >16ms / 230VAC, >16ms / 115VAC at full load

Input
Voltage Range: 90 ~ 264VAC, 127 ~ 373VDC
Frequency Range: 47 ~ 63Hz
Efficiency (Typ.) at 230VAC: 90% 90% 91% 91%
AC Current (Typ.): 1.71A / 100VAC, 1.23A / 240VAC
Inrush Current (Typ.): Cold Start 30A / 115VAC, 60A / 230VAC
Leakage Current: For earth For earth < 275uA / 264VAC, For patient < 100uA / 264VAC

Protection

Over Load: Above 110% rated output power Protection type: latch-style (Recovery after reset AC power ON)
Over Voltage: 115 ~ 140% rated output voltage Protection type: latch-style (Recovery after reset AC power ON)

Environment
Working Temp.: -20 ~ +70°C (Refer to output load de-rating curve)
Working Humidity: 20 ~ 90% RH non-condensing
Storage Temp. & Humidity: -40 ~ +85°C, 10 ~ 95% RH
Temp. Coefficient: ±0.03% / °C (0 ~ 50°C)
Vibration: 10 ~ 500Hz, 2G 10min. / 1cycle, period for 60min. each along X, Y, Z axes

Safety & EMC
Safety Standards: Meets UL / cUL (ANSI / AAMI ES 60601-1; CAN / CSA-C22.2 No.60601-1); NFPA 99
Meets TUV (EN 60601-1; IEC 60601-1); MOPP approval
Withstand Voltage: I/P-O/P: 4KVAC, I/P-FG: 1.8KVAC, O/P-FG: 1.5KVAC
Isolation Resistance: I/P-O/P: 100M Ohms / 500VDC
EMI Conduction & Radiation: Certified EN 55011
Harmonic Current: Certified EN 61000-3-2; EN 61000-3-3
EMS Immunity: Certified EN 60601-1-2; IEC 61000-4-2, 3, 4, 5, 6, 8, 11

Others
MTBF: XXX K HRS Certified MIL-HDBK-217F
Cooling: Free Air convection for 150W, with 23.5 CFM Fan for 250W
Dimension (WxHxH): 50.8x32.1x101.6 mm / 2.00x1.26x4.00 inch
Packing: 0.32 kg

Note
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25°C of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uF & 47uF parallel capacitor.
3. Tolerance: includes set up tolerance, line regulation and load regulation.
4. Line regulation is measured from low line to high line at rated load.
5. Load regulation is measured from 0% to 100% rated load.
6. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives.
Derating CurveUser-added image

 
Qty (ea.)/CartonTBD
Individual Box/Gift Box Size (mm)130*77*45
Individual Box/Gift Box ColorWhite
Master Carton Size (mm)TBD
Gross Weight (kg)TBD
Label Width (mm)TBD
Label Height (mm)TBD
Input Configuration
Description3 Pin connector Mating Housing: JST VHR-3N or equivalent Mating Contact: JST SVH-21T-P1.1 or equivalentUser-added image
Approvals

Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.

Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.

LogoDescription
User-added imageAttention Mark / Refer to operating manual for operating and service instructions
User-added imageUKCA Certification
User-added imageJapan: Voluntary Control Council for Interference (VCCI)
 User-added imageWEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm) Mark is on the label or Molded in the case
User-added imageCHINA SJ/T 11364-2014, China RoHS Chart: http://www.globtek.com/pdf/F-GT-DJD-8.4.1-006%20China%20RoHS%20Declaration%205-20-22.pdf
User-added imageUkraine UKRSepro (Document: www.globtek.com/html/iso_certificates/GT_Ukraine.pdf)
User-added image
 
GOST-R mark for Russia(Document: http://www.globtek.com/html/iso_certificates/GT_GOST-R.pdf). 03/15/2015. Due to lack of demand, GlobTek will not be renewing it's GOST certification. Please contact your sales engineer for pricing on EAC.

User-added image  
CE Certification
User-added imageJAPAN TUV Rheinland-PSE GlobTek Inc to J60950-1(H26) , J55022(H22),J3000(H25).Please follow the procedure listed in the following link for proper import to Japan: https://www.globtek.com/r2/Szj4Vb
User-added imageUR 60601-1, 3rd Edition, CAN/CSA-C22.2 No. 601.1-1
User-added imageUL60950-1, cULus to UL 60950-1 E170507
User-added imageIndoor Use Only - Mark is on the label or Molded in the case
RoHSSpecifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf
User-added imagenull
Request Info