Model NumberGTM43007-CWWVV-F
DescriptionGTM43007-CWWVV-F, ITE / Medical Power Supply, 60601-1-4th Ed. , Open Frame/Internal, Regulated Switchmode AC-DC, , Input Rating: 100-240V~, 50-60 Hz, Molex 26-60-4030 3 Position Header Pin 1: Line, Pin 2: Removed, PIN 3: Neutral, Output Rating: 60 Watts, 5-48V in 0.1V increments, Approvals: CB 60601-1 2MOPP CE China RoHS Class I RoHS 2 Ukraine WEEE CE FCC CB CB EN/IEC 60335-1 cETLus ES60601 cETLus cETLus 60601-1 3rd Double Insulation Level V LPS NEMKO EN/IEC 60335-1 WDaE S-Mark IEC/EN 60601-1
Agency Documentshttp://www.globtek.info/certs/GTM43007/
CE EC-Declaration for Medicalhttp://globtek.force.com/models/ec_declaration?id=a0Oa000000Mgc7lEAB
CE EC-Declaration for Industrial/Commercial/ITEhttp://globtek.force.com/models/ec_declaration?id=a0Oa000000Mgc7lEAB
RoHS/RoHS2 Declarationhttp://www.globtek.com/pdf/rohs_cert/a0Oa000000Mgc7lEAB
REACH Declarationhttp://www.globtek.com/pdf/iso_certificates/REACH.pdf
Conflict Minerals Declarationhttp://www.globtek.com/pdf/conflict-minerals.pdf
Model Parameters
TypeOpen Frame/Internal
TechnologyRegulated Switchmode AC-DC
CategoryITE / Medical Power Supply
Input Voltage100-240V~, 50-60 Hz
I/P Amps (A)1.5A
Wattage (W)60
Vout Range (V)5-48
EfficiencyN/A
Ingress ProtectionN/A
Size (mm)127x76.2x22.5
Enclosure
User-added image




















 
Rating Table
Model NumberVoltageAmps (A)Watts (W)RFQ
GTM43007-C3005-F5630RFQ
GTM43007-C3007-1.0-F6530RFQ
GTM43007-C3007-F74.2829.96RFQ
GTM43007-C3809-15-F7.5537.5RFQ
GTM43007-C4509-F9545RFQ
GTM43007-C4512-2.0-F104.545RFQ
GTM43007-C4512-F123.7545RFQ
GTM43007-C6015-F15460RFQ
GTM43007-C6018-F183.3359.94RFQ
GTM43007-C6024-5.0-F193.1559.85RFQ
GTM43007-C6024-4.0-F20360RFQ
GTM43007-C6024-F242.560RFQ
GTM43007-C6036-6.0-F30260RFQ
GTM43007-C6036-F361.6659.76RFQ
GTM43007-C6048-8.0-F401.560RFQ
GTM43007-C6048-F481.2560RFQ
SpecificationsA) ELECTRICAL SPECIFICATIONS:
01. Input Voltage: Specified 90-264 Vac, Nameplate rated: 100-240Vac
02 .Input Frequency: Specified 47-63 Hz, Nameplate rated 50-60Hz
03. Output Regulation: +/- 5% measured at the output connector
04. Line Voltage Regulation: +/- 1% typical measures at full load
05. Output Ripple (Vp-p): 1% or 150 mV whichever is greater measured at 20 MHz bandwidth with 0.1 uf ceramic capacitor in parallel with 10 uf electrolytic capacitor connected at the end of output connector at nominal line
06. Turn-On/Turn-Off Overshoot: 5% maximum, 1mS typical recovery time for 25% step load
07. Turn-On Delay: 3 seconds typical
08. Hold-Up Time: 8mS typical @ nominal input voltage & full load
09. Inrush Current: 30A typical @ 115Vac input ; 60A typical @ 230Vac input
10. Switching Frequency: 66.5 KHz typical

B) PROTECTION
1. Over-Voltage: Protected via zener diode across the output
2. Short Circuit: Electronically Protected, unit will recover upon removal of fault
3. Input Protection: Input dual-fusing (for both line and neutral)

C) SAFETY
1. Dielectric Withstand Voltage: 5656Vdc from primary to secondary
2. Earth Leakage Current: Did not exceed 5 mA in NORMAL CONDITION and 10 mA in SINGLE FAULT CONDITION
3. Touch Current: Did not exceed 100uA NC(Normal condition) 500uA SFC(single fault condition)
4. Means of Protection: 2 x MOPP (Means Of Patient Protection)

D) OTHER:
1. MTBF: 200,000 hours @ 25°C ambient temperature
2. Operating Temperature: -10°C to 45°C ambient temperature, Convection Cooled. Higher ambient temperatures available with forced air cooling.
3. Humidity: 0% to 93% relative humidity, passed 7 day humidity testing per IEC 60601-1
4. Storage Temperature: -10°C to 80°C
5. RoHS 2: Complies with EU 2011/65/EU and China SJ/T 11363-2006

E) OPEN FRAME
01. Markings: Label
Derating CurveUser-added image
Qty (ea.)/Carton60
Individual Box/Gift Box Size (mm)147*100*33
Individual Box/Gift Box ColorWhite
Master Carton Size (mm)455*415*210
Gross Weight (kg)15.2
Label Width (mm)
Label Height (mm)
Input Configuration
DescriptionMolex 26-60-4030 3 Position Header Pin 1: Line, Pin 2: Removed, PIN 3: Neutral
User-added image     
                                                     User-added image
    
Approvals
LogoDescription
User-added imageBook 60335 up to 36 Volts
User-added image60601-1 3rd Operating Instructions
No Logo ApplicableCB report IEC60601-1 2005 A1+C1+C2 2016-2-4 and or EN 60601-1:2006 3.1rd Edition 2xMOPP
No Logo CB Report EN60950-1

No Logo Applicable

CB to 60335-1 5th Edition (Safety of Household and Similar Electrical Appliances, Part 1: General Requirements) CB Report/ Certificate IEC 60335-1:2010 Up to 36 volts only
User-added image  CE Certification CE Mark: tested to comply with EN 55032.2012+AC.2013 EN 61000-3-2.2014 EN 61000-3-3.2013 EN 60601-1-2.2015 EN 55024.2010
User-added image  CE Mark: tested to comply with EN 60601-1-2:2007 Medical electrical equipment Part 1-2: - Mark is on the label or Molded in the case
Conforms to ANSI/AAMI ES 60601-1:2005
Cert.to CAN/CSA Std. C22.2
 
cETLus ANSI_AAMI ES60601-1_2005
User-added imageConforms to ANSI/UL Std. 60950-1 2nd Edition Cert. to CAN/CSA Std.C22.2 NO.60950-1 2nd Edition
User-added image
 
cETLus-Cert-4007497 ANSI/AAMI ES60601-1:2005/A1:2012 Issued: 2012/08/20 Medical electrical equipment – Part 1: Amendment 1; CAN/CSAC22.2 No. 60601-1:14, Third Edition Issued: 2014/03/01 - Medical Electrical Equipment - Part 1
User-added imageCHINA SJ/T 11364-2014, China RoHS Chart: http://en.globtek.com/globtek-rohs.php
User-added image
Class I Earth Ground
User-added image

Double Insulation - Mark is on the label or Molded in the case
User-added image
FCC PART 15 CLASS B
User-added imageFuse 60335 T2.0A 250VAC Up to 36 volts
User-added imageIndoor Use Only - Mark is on the label or Molded in the case
 User-added image

Efficiency: complies to section 301 of Energy Independence and Security Act (EISA) complies with Energy Star tier 2 (North America), ECP tier 2 (China), MEPS tier 2 (Australia), Code of Conduct (Europe)
LPSLimited Power Source
User-added imageNEMKO IEC/EN 60335-1 (Safety of Household and Similar Electrical Appliances, Part 1: General Requirements) up to 36 Volts
RoHS 2RoHS 2: Complies with EU 2011/65/EU and CHINA SJ/T 11363-2006 http://www.ce-mark.com/Rohs%20final.pdf
User-added imageSemko S-Mark-Cert-EN60601-1 and EN60950-1
User-added imageUkraine UKRSepro (Document: www.globtek.com/html/iso_certificates/GT_Ukraine.pdf)
WDaEAppliances can be used in warm damp equable climates
 User-added imageWEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm) Mark is on the label or Molded in the case
News Releases
TitleWeb Link
Open Frame ITE & Medical 60W Switching Power Supply Offers superior Isolation and high performance with a universal input of 90-264VDC and 5-48VDC Output in both Class I and Class II configurations, Model GTM43007http://en.globtek.com/news/open-frame-ite-medical-60w-switching-power-supply-offers-superior-isolation-and-high-performance-with-a-universal-input-of-90-264vdc-and-5-48vdc-output-in-both-class-i-and-class-ii-configurations-model-gtm43007

Begär information