World map Svenska

ICT / ITE / Medical Power Supply/Class 2/Household Power Supply

Information
Model NumberGTM96600-65VV.V-TP(SR with fixed Plug 2 pin)
DescriptionGTM96600-65VV.V-TP(SR with fixed Plug 2 pin), ICT / ITE / Medical Power Supply/Class 2/Household Power Supply, 60601-1-4th Ed. , Desktop/External, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60Hz, See "Blade/Cord Installed" section. Contact GlobTek for customization requests such as Molded country plugs, Molex or TE connectors, or IEC connectors mounted on any type of cable in any customer specified lengths., Output Rating: 65 Watts, Power rating with convection cooling (W) , 5-54V in 0.1V increments, Approvals: IP42; Morocco; CB 62368; S-Mark IEC/EN 60601-1; CB 60601-1; CE; China RoHS; Double Insulation; cETLus 60601-1; FCC; PSE; PSE; PSE; Level VI; RoHS; Ukraine; VCCI; WEEE; EAC; UKCA; cETLus 62368-1;
Agency Documentshttp://www.globtek.info/certs/GTM96600-TZ/
CE EC-Declaration for Medicalhttps://globtek.secure.force.com/models/ec_declaration?id=a0O3a00000LAahgEAD
CE EC-Declaration for Industrial/Commercial/ITEhttps://globtek.secure.force.com/models/ec_declaration?id=a0O3a00000LAahgEAD
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for Medicalhttps://globtek.secure.force.com/models/ukca_declaration?id=a0O3a00000LAahgEAD
UKCA-Declaration for Industrial/Commercial/ITEhttps://globtek.secure.force.com/models/ukca_declaration?id=a0O3a00000LAahgEAD
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://globtek.secure.force.com/models/Morocco_SDoC_Declaration?id=a0O3a00000LAahgEAD
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0O3a00000LAahgEAD
Prop 65 Statementhttps://en.globtek.com/prop-65-statement.pdf
REACH Declarationhttps://www.globtek.com/pdf/iso_certificates/REACH.pdf
Conflict Minerals Declarationhttps://www.globtek.com/pdf/conflict-minerals.pdf
Model Parameters
TypeDesktop/External
TechnologyRegulated Switchmode AC-DC Power Supply AC Adaptor
CategoryICT / ITE / Medical Power Supply/Class 2/Household Power Supply
Input Voltage Range (V)100-240V~, 50-60Hz
Max Input Current (A)1.5A
Max Output Power Rating (W)65
Output Voltage Range (V)5-54
EfficiencyUSA DOE Level VI / Eco-design Directive 2009/125/EC, (EU) 2019/1782
Ingress ProtectionIP42
Size (mm)117.5*53.5*34.2
EnclosureUser-added image
Rating Table
Model Number (Final ordering #/Part No to be defined) VoltageAmps (A)Watts (W) RFQ
GTM96600-4005-TP* 5840
GTM96600-4005.95-TP* 5.956.7239.98
GTM96600-4808-TP* 8648
GTM96600-5409-TP* 9654
GTM96600-6010-TP* 10660
GTM96600-6011-TP* 115.4559.95
GTM96600-6512-TP* 125.4164.92
GTM96600-6513-TP* 13565
GTM96600-6514-TP* 144.64264.99
GTM96600-6515-TP* 154.3364.95
GTM96600-6516-TP* 164.06264.99
GTM96600-6517-TP* 173.82364.99
GTM96600-6518-TP* 183.6164.98
GTM96600-6519-TP* 193.42165
GTM96600-6520-TP* 203.2565
GTM96600-6524-TP* 242.764.8
GTM96600-6527-TP* 272.40764.99
GTM96600-6530-TP* 302.16664.98
GTM96600-6535-TP* 351.85765
GTM96600-6548-TP* 481.3564.8
* (Final ordering #/Part No to be defined)
Specifications

Input

Input Voltage:Specified: 90-264VAC, Nameplate: 100-240VAC
     100% rated load current for 90-264VAC
       85% rated load current for 85-264VAC
     100% rated load current for 110-370VDC
Input Frequency:Specified: 47-63Hz, Nameplate: 50-60Hz
No Load Input Power:<75mW @ 230VAC (EU CoC Tier 2 compliant) for ≤ 49W rating
<150mW @ 230VAC (EU CoC Tier 2 compliant) for > 49W rating
Inrush Current:30A/60A max. (cold start @ 115V/230VAC)
Efficiency:DoE Level VI and EU CoC Tier 2 compliant

Output

Turn-on Delay:1 second max. @ 115VAC and full load
Output Regulation:± 5% (measured at output connector)
Line Regulation:± 0.5% typ. (measured at output connector)
Ripple:1% or 100mV, whichever is greater (using a 47µF low-ESR cap + 0.1µF ceramic capacitor, measured @ 20MHz BW, at output connector)
Transient Response:5% max. deviation, 1ms typ. recovery time (with 40% to 70% load step)
Hold-up Time:8ms typ. (nominal input voltage and full load)
Power Indicator:Green LED

Protections

Input Protection:MOV transient suppressor, input line fusing
Over-Current Protection:110 - 160% of rated output current, auto-restart, current limit range determined by normal output power rating, not de-rated power rating
Short-Circuit Protection:Auto-restart
Over-Voltage Protection:110 - 130%, latched off, cycle AC to reset
Over-Temperature Protection:On-chip temperature sensor (integrated in primary-side switching controller), shut down at Tj = ~140°C, latched off, cycle AC to reset

Environmental

MTBF:1,500,000 hours @ 25°C ambient, full load (Telcordia SR-332, Issue 3)
Operating Temperature:-20°C to 40°C (full load)
Storage Temperature:-30°C to 80°C
Humidity:0% to 95% relative humidity, non-condensing
Altitude:5000m
Cooling:Convection
RoHS:Complies with EU 2011/65/EU and China SJ/T 11363-2006
Note 1:Regarding operating temperature
a. See below de-rating table for output power capability at alternate temperature
b. Extended low-end temperature range available as a custom option

Safety

Dielectric Withstand Voltage:4000VAC or 5656VDC from input to output
3000VAC or 4242VDC from input to earth (Class I models only)
Touch Current:Class I: 20µA max.
Class II: 85µA max.
Class II F2: 20µA max.
Earth Leakage CurrentClass I: 300µA max.
Class II: N/A
Means of Protection:2 x MOPP
Output Isolation Options:Class I: C6 or C14 inlet (output negative tied to Earth contact)
Class II: C8 or C18 inlet
Class II FE: C6 or C14 inlet (output isolated from Earth contact)
Review isolation options by reading our white paper PSU Isolation and Identity.
Earth Continuity Test:< 0.1Ω between earth pin (on AC inlet) and PCB termination point (Class I models only)
Compliant Standards:See listing at end of this datasheet for specific standards

EMC

Applicable Standards:Medical: EN60601-1-2, 4th edition
Emissions: EN55032, EN61000-6-3, EN61000-6-4, CISPR11, CISPR22
Immunity: EN55024, EN61000-6-1, EN61000-6-2
Conducted Emissions:Class B, FCC Part 15, Class B (with resistive load)
Radiated Emissions:Class B, FCC Part 15, Class B (with resistive load)
Harmonic Current Voltage Distortion:EN61000-3-2, Class A
Voltage Fluctuations/Flicker:EN61000-3-3
Electrostatic Discharge (ESD) Immunity:EN61000-4-2, 10KV contact discharge, 20KV air discharge
Radiated RF Immunity:EN61000-4-3, 10V/m 80-1000MHz,  3V/m 1-2.7GHz, 80% 1KHz AM
EFT/Burst Immunity:EN61000-4-4, 4KV/100kHz.
Line Surge Immunity:EN61000-4-5, 2KV differential, 4KV common-mode
Conducted RF Immunity:EN61000-4-6, 3VRMS, 80% 1KHz AM
Power Frequency Magnetic Field Immunity:EN61000-4-8, 3A/m
Voltage Dip Immunity:EN61000-4-11

Enclosure

Housing:High impact plastic, 94V0 polycarbonate, non-vented
AC Input Mechanical Options:Desktop: C6, C7, C14, or C18 inlet
Hybrid: Desktop inlet or changeable blades for wall-plug-in style, Class I or Class II input
Markings:Label and/or pad printed and/or laser-etched into case

Special Options

  1. Removal of LED (cost reduction or aesthetic purpose)
  2. Custom cordsets, various cordage types, and connector types
  3. Custom markings
  4. Short term output surge capability for motors or other high peak current loads. (Available up to 120W for 1 second at nameplate voltage range. Higher levels possible for limited input voltage range, or larger allowed voltage dip.)
  5. Reduced leakage current
  6. Tightened output voltage tolerance
  7. Reduced output ripple level
  8. Reduced output power rating
  9. High reliability PCB laminate with plated through-holes for IPC610 Class 2 compliance
  10. Low temperature option, down to -40°C
  11. Special housing and cordset colors
  12.  "Back EMF" upgrade, for motor loads and other high inductance applications with reverse energy flow requirements
  13. Improved ingress protection (IP) rating

Output Connectors

 

 

Derating CurveUser-added image
+74C operation with an output power of 19.2W; and cold start at -40C
 
Qty (ea.)/Carton20
Individual Box/Gift Box Size (mm)Egg Crate
Individual Box/Gift Box Color
Master Carton Size (mm)427 x 405 x 158
Gross Weight (kg)15.6
Label Width (mm)35.0
Label Height (mm)70.0
Input Configuration
DescriptionSee "Blade/Cord Installed" section. Contact GlobTek for customization requests such as Molded country plugs, Molex or TE connectors, or IEC connectors mounted on any type of cable in any customer specified lengths.
Output Configuration
Common output connector options:
Click to add to your request.
L Type (Coaxial 5.5x2.5mm plug)
C Type (Coaxial 5.5x2.1mm plug)
K Type (Coaxial 3.5x1.3mm plug)
LL Type (5.5x2.5mm Locking 760k type)
CL Type (5.5x2.1mm Locking S761k type)
Magnetic Connector
L Type (Coaxial 5.5x2.5mm plug) C Type (Coaxial 5.5x2.1mm plug) K Type (Coaxial 3.5x1.3mm plug) LL Type (5.5x2.5mm Locking 760k type) CL Type (5.5x2.1mm Locking S761k type) Magnetic Connector
YL3 Type (KPPX-3P)
YL4 Type (KPPX-4P)
EJ1/2/3/4/5 (EIAJ RC-5320A type connectors)
MSB Type (Micro USB)
Custom design
YL3 Type (KPPX-3P) YL4 Type (KPPX-4P) EJ1/2/3/4/5 (EIAJ RC-5320A type connectors) MSB Type (Micro USB) Inquire for custom design

For a comprehensive list of options, click here

Contact GlobTek for your specific requirements or custom solutions.

Approvals
LogoDescription
Medical 4th Ed.EN/IEC 60601-1-2, 4th Edition (Please visit EMC reports for further info)

No Logo Applicable

CB for IEC 60601-1:2005+A1

No Logo Applicable

CB for IEC 62368-1:2014 (Second Edition)
User-added image  Test standard: EN 55032:2012+AC:2013 EN 61000-3-2:2014 EN 61000-3-3:2013 EN 55024:2010 EN 60601-1-2:2015
User-added imageCHINA SJ/T 11364-2014, China RoHS Chart: http://en.globtek.com/globtek-rohs.php
Conforms to UL STD. 62368-1
Certified to CSA STD C22.2 NO.62368-1
Conforms to UL STD. 62368-1 Certified to CSA STD C22.2 NO.62368-1
User-added image
User-added image  Declaration ДС № ЕАЭС N RU Д-US.КА01.В.10453_19 Custom Union of Russia, Belarus and Kazakhstan http://www.globtek.com/redirect/?loc=gost-certificate-eac-declaration
User-added imageMedical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance [AAMI ES60601-1:2005 +A1]
User-added imageCompliance of this power supply with FCC Part 15, Class B has been demonstrated with a standard output load. The FCC law stipulates that system-level testing is required to demonstrate compliance with the FCC emission limits with the actual system load.
User-added imageIndoor Use Only - Mark is on the label or Molded in the case
IP42Ingress Protection: ?IP42 to IEC60529:2001 Protection against granular foreign bodies - Protected against direct sprays up to 15o from the vertical Optional
User-added imageJAPAN TUV R-PSE, Cert. No. JD 50313287, to J60950-1(H26) , J55022(H22),J3000(H25)[DC15?30V]. Please reference the following website for guidelines on PSE regulations: http://en.globtek.com/importing-ite-and-medical-power-supplies-ac-adaptors-to-japan/
User-added imageJAPAN TUV R-PSE, Cert. No. JD 50313287, to J60950-1(H26) , J55022(H22),J3000(H25)[DC30?60V]. Please reference the following website for guidelines on PSE regulations: http://en.globtek.com/importing-ite-and-medical-power-supplies-ac-adaptors-to-japan/
User-added imageJAPAN TUV R-PSE, Cert. No. JD 50469658, to J62368-1(H30) , J55032(H29),J3000(H25)[15V or less]. Please reference the following website for guidelines on PSE regulations: http://en.globtek.com/importing-ite-and-medical-power-supplies-ac-adaptors-to-japan/
 User-added image
 
Efficiency: complies to section 301 of Energy Independence and Security Act (EISA) complies with Energy Star tier 2 (North America), ECP tier 2 (China), MEPS tier 2 (Australia), Code of Conduct (Europe)
User-added image
 
Morocco SDoC declaration http://www.globtek.info/certs/Morocco%20SDoC%20Declaration/
RoHSSpecifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf
User-added image
 
S-Mark Certificate EN 60601-1:2006+A1+A12, EN 60601-1-11:2015
User-added image
Serial Number Information
User-added imageUKCA Certification
User-added imageUkraine UKRSepro (Document: www.globtek.com/html/iso_certificates/GT_Ukraine.pdf)
User-added imageJapan: Voluntary Control Council for Interference (VCCI)
 User-added imageWEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm) Mark is on the label or Molded in the case
Popular Products (Stock items)
SKUVolts OutCurrent Out (Amps)ConnectorRFQSample Request
Vad är ett externt AC/DC adapter ? externa desktop AC/DC adaptrar är ochså känd som AC adaptrar eller AC/DC omvandlare och är en typ av externa strömförsörjningar. Vissa elektriska apparater kräver ström, men saknar interna komponenter för att få fram önskad spänning och ström. Det är här ett AC adapter användas för att omvandla AC ström till DC spänning. Vad är ett IEC 62368-1 : 2018 power supply ? IEC 62368-1 : 2018 användas till säkerhet för elektriska och elektroniska apparater angående audio, video, information och kommunikationsteknologi, business- och kontormaskiner med en graderad spänning som ej överstiger 600V. Detta är en produktsäkerhetsstandard som klassifiserar energikällor, föreskrivna skydd mot dessa energikällor och ger information vid användning av- och krav till dessa skydd. Föreskrivna skyddskrav är menade för att minska sannolikhet för smärta, injurier och i fall av brand och egendom skador. Et IEC 62368 godkänd AC/DC power supply, är ett adapter som är testad och auktoriserad enligt bestämda standard. Ersätter ett IEC 62369-1 godkänd power supply ett IEC 60950-1 godkänd power supply ? 20 december 2020 är säkerhets- standardene IEC 60950-1 och IEC 60065 historia och ersättes av IEC 62368-1angående tekniska - och audio video produkter. I nord-Amerika vill ETL och UL 60950-1 godkanningar fortfarande existera och vill antagligen användas ytterligare. Vad är Level VI effektivitet ? Nummer I till VI är en international effektmärkeprotokoll som tillåter producenter att designa effektivitetsnivået av ett produkt. Detta är baserad på mätningar av energieffektiviteten och konsumtion vid no load av externa AC/DC adaptrar eller power supply som testas vid olika lastnivåer. No Load konsumptionen refererar till konsumption som ett externt power supply konsumerar även om produktet som önskas strömförsörjad är avstängd eller koplat av. Är Level VI ett krav ? Level VI effektivitet är obligatorisk i många länder och regioner för särskilda produkter och applikationer. Vad är ECO Design Directive 2009/125/EC ? ECO . Design Direktivet är lagstadgad i Europa i dom senaste energikrav för bevarande av AC/DC adaptrar och externa power. Hur fungerar ett AC/DC adapter ? AC/DC adapteren eller exsternt power supply är bestående av ett power supply som är skyddad i ett plasthus motsvarande en AC plugg, därmed namn som plug type, wall plug-in , plug in eller plugtop adapter. Adapteren monteras i vägguttaget och
Begär information

Model

Call
Email
💬
Text
💁
Chat
?
Begär en offert
Kan vi hjälpa dig?
Kan vi hjälpa dig?