ICT / ITE / Medical Power Supply

Information
Model NumberGTM43007-AWWVV-F (D3 with standoffs)
DescriptionGTM43007-AWWVV-F (D3 with standoffs), ICT / ITE / Medical Power Supply, 60601-1-4th Ed. , Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, Solder Pins for Line and Neutral, 1.14 mm square pins, 11.43 mm long, Output Rating: 60 Watts, Power rating with convection cooling (W) , 5-48V in 0.1V increments, Approvals: CB EN/IEC 60335-1; EAC; ETL 62368; CE; Fuse 60335; WDaE; Level V; S-Mark IEC/EN 60601-1 60950; NEMKO EN/IEC 60335-1; cETLus 60601-1 3rd; cETLus; cETLus ES60601; FCC; CE; WEEE; Ukraine; RoHS; Class I; China RoHS; CB 62368; CB 60601-1 2MOPP; Book 60601; Fuse 60335; CB 60950; cETLus 62368-1; UKCA; CCC; GTM43007-AWWVV-F (D3 with standoffs), ICT / ITE / Medical Power Supply, 60601-1-4th Ed. , Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, Solder Pins for Line and Neutral, 1.14 mm square pins, 11.43 mm long, Output Rating: 60 Watts, Power rating with convection cooling (W) , 5-48V in 0.1V increments, Approvals: CB EN/IEC 60335-1; EAC; ETL 62368; CE; Fuse 60335; WDaE; Level V; S-Mark IEC/EN 60601-1 60950; NEMKO EN/IEC 60335-1; cETLus 60601-1 3rd; cETLus; cETLus ES60601; FCC; CE; WEEE; Ukraine; RoHS; Class I; China RoHS; CB 62368; CB 60601-1 2MOPP; Book 60601; Fuse 60335; CB 60950; cETLus 62368-1; UKCA; CCC;
Agency Documentshttp://www.globtek.info/certs/GTM43007/
CE EC-Declaration for Medicalhttps://globtek.secure.force.com/models/ec_declaration?id=a0O0c00000PIXPlEAP
CE EC-Declaration for Industrial/Commercial/ITEhttps://globtek.secure.force.com/models/ec_declaration?id=a0O0c00000PIXPlEAP
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for Medicalhttps://globtek.secure.force.com/models/ukca_declaration?id=a0O0c00000PIXPlEAP
UKCA-Declaration for Industrial/Commercial/ITEhttps://globtek.secure.force.com/models/ukca_declaration?id=a0O0c00000PIXPlEAP
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://globtek.secure.force.com/models/Morocco_SDoC_Declaration?id=a0O0c00000PIXPlEAP
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0O0c00000PIXPlEAP
Prop 65 Statementhttps://en.globtek.com/prop-65-statement.pdf
REACH Declarationhttps://www.globtek.com/pdf/iso_certificates/REACH.pdf
Conflict Minerals Declarationhttps://www.globtek.com/pdf/conflict-minerals.pdf
Model Parameters
TypeOpen Frame/Internal
TechnologyRegulated Switchmode AC-DC Power Supply AC Adaptor
CategoryICT / ITE / Medical Power Supply
Input Voltage Range (V)100-240V~, 50-60 Hz
Max Input Current (A)1.5A
Max Output Power Rating (W)60
Output Voltage Range (V)5-48
EfficiencyN/A
Ingress ProtectionN/A
Size (mm)76.2x50.8x23.6 for 5V-14V models, 76.2x50.8x22.5 for 15V-48V modelsTop Diagram i
EnclosureUser-added image
Top Diagram is for 5V-14V models, Bottom Diagram is for 15V-48V models
User-added image

INPUT CONNECTOR for POWER
User-added image
INPUT CONNECTOR for GND
User-added image
 
Rating Table
Model Number (Final ordering #/Part No to be defined)VoltageAmps (A)Watts (W)RFQ
* (Final ordering #/Part No to be defined)
Specifications
A) ELECTRICAL SPECIFICATIONS:
01. Input Voltage: Specified 90-264 Vac, Nameplate rated: 100-240Vac
02. Input Frequency: Specified 47-63 Hz, Nameplate rated 50-60Hz
03. Output Regulation: +/- 3% measured at the output connector
04. Line Voltage Regulation: +/- 0.5% maximum measured at full load
05. Output Ripple (Vp-p): 1% or 150 mV whichever is greater, measured at 20 MHz bandwidth with 0.1 uf ceramic capacitor in parallel with a low impedance 47 uf electrolytic capacitor connected 3” from the output connector .
06. Turn-ON/OFF: 5% maximum,1 mS typical recovery time for 25% step load
07. Turn-ON Delay: 3 Sec max
08. Hold-Up Time:8 mS typical @ nominal input voltage and full load
09. Inrush Current: 30A maximum  @ 115 Vac input, 60A maximum  @ 230 Vac input
10. Switching Frequency: 65 KHz typical
11. Efficiency: 87% @115Vac, 89% @ 230Vac typical at full load
12. No Load Power: < 300mW @ 230Vac

B) PROTECTION
01. Overvoltage Protection: 110% to 130% of nominal output voltage, and less than 60Vdc
02. Short Circuit/Over-load: momentary short circuit protected, output will auto recover upon removal of short
03. Input Protection: Input line fusing and 300Vac MOV

C) SAFETY
01. Dielectric Withstand Voltage: 4000Vac / 5656Vdc from primary to secondary
02.  Output Isolation Options:  Class I or Class II
03. Earth Leakage Current: Does not exceed 300uA in NORMAL CONDITION and 1000uA in SINGLE FAULT CONDITION
04. Output Leakage Current: Does not exceed 75uA NC(Normal condition) 300uA SFC(single fault condition)
05. Means of Protection: 2 x MOPP
06. Compliant Standards: See listings at end of this drawing for specifics

D) EMC
EN 60601-1-2, 4th edition
Emissions, per EN 55032, EN 61000-6-3, EN 61000-6-4, CISPR11 and CISPR22
  Conducted Emissions: Class B, FCC Part 15, Class B
  Radiated Emissions: Class B, FCC Part 15, Class B
Line Frequency Harmonics EN61000-3-2, Class A
Voltage Fluctuations/Flicker EN61000-3-3
 
Immunity, per EN 55024, EN 61000-6-1, EN 61000-6-2
  Static Discharge Immunity EN61000-4-2, 10kV Contact Discharge, 20kV air discharge
  Radiated RF Immunity EN61000-4-3, 10V/m 80-1000MHz, 3V/m 1-2.7GHz, 80% 1KHz AM.
  EFT/Burst Immunity EN61000-4-4, 4kV/100kHz.
  Line Surge Immunity EN61000-4-5, 2kV differential, 4kV common-mode
  Conducted RF Immunity EN61000-4-6, 3Vrms, 80% 1KHz AM
  Power Frequency Magnetic Field Immunity EN61000-4-8, 3A/m
  Voltage Dip Immunity EN61000-4-11, Criteria

E) OTHER
01. MTBF: 1,000,000 Hours @ 45°C ambient temperature, Full Load
02. Operating Temperature: -10°C to 45°C ambient temperature
03. Humidity: 0% to 95% relative humidity
04. Storage Temperature: -30°C to 80°C
05. Cooling: Convection
06: Elevation: 5000M
07. ROHS 2: Complies with EU 2011/65/EU and China SJ/T 11363-2006


F) MECHANICAL  FORMAT
01. Type: Open Frame
02. Size: A suffix version, (2”x3”), 76.2 x 50.8 x 30.5 mm
03. Size: B suffix version, (2”x4”), 101.6 x 50.8 x 30.5 mm
04. Size: C suffix version, (3”x5”), 127.0 x 76.2 x 30.5 mm


G) SPECIAL OPTIONS
01. Custom Label Markings
02. Short term Output Surge Capability
03. Reduced Leakage Current versions
04. Tightened output voltage tolerance
05. Reduced output ripple level
06. High Rel PCB laminate with Plated through Holes for IPC610 Class 2 Compliance
07. Low Temperature operation, down to -40°C
08. Quasi Constant Current Output, for Lead-Acid Battery Charger Applications
09. Special Input or Output Connectors
10. Over-Temperature Protection, Latched Shut-Down
 
Derating CurveUser-added image
 
Qty (ea.)/Carton124
Individual Box/Gift Box Size (mm)95*65*35
Individual Box/Gift Box ColorWhite
Master Carton Size (mm)442 x 330 x 275
Gross Weight (kg)15.2
Label Width (mm)
Label Height (mm)
Input Configuration
DescriptionSolder Pins for Line and Neutral, 1.14 mm square pins, 11.43 mm long
Approvals

Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.

Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.

LogoDescription
Medical 4th Ed.EN/IEC 60601-1-2, 4th Edition (Please visit EMC reports for further info)
User-added imageBook 60335 up to 36 Volts
User-added image60601-1 3rd Operating Instructions
No Logo ApplicableCB report IEC60601-1 2005 A1+C1+C2 2016-2-4 and or EN 60601-1:2006 3.1rd Edition 2xMOPP
No Logo CB Report EN60950-1

No Logo Applicable

CB for IEC 62368-1:2014 (Second Edition)

No Logo Applicable

CB to 60335-1 5th Edition (Safety of Household and Similar Electrical Appliances, Part 1: General Requirements) CB Report/ Certificate IEC 60335-1:2010 Up to 36 volts only
User-added imageCCC Altitude up to 5000 m GB17625.1-2012, GB4943.1-2011, GB/T9254-2008 for GT*43007
User-added image  CE Certification CE Mark: tested to comply with EN 55032.2012+AC.2013 EN 61000-3-2.2014 EN 61000-3-3.2013 EN 60601-1-2.2015 EN 55024.2010
User-added image  CE Mark: tested to comply with EN 60601-1-2:2015 Medical electrical equipment Part 1-2: - Mark is on the label or Molded in the case
User-added imageAudio/Video, Information And Communication Technology Equipment - Part 1: Safety Requirements [UL 62368-1:2014 Ed.2] Audio/Video, Information And Communication Technology Equipment - Part 1: Safety Requirements [CSA C22.2#62368-1:2014 Ed.2]
Conforms to ANSI/AAMI ES 60601-1:2005
Cert.to CAN/CSA Std. C22.2
 
cETLus ANSI_AAMI ES60601-1_2005
User-added imageConforms to ANSI/UL Std. 60950-1 2nd Edition Cert. to CAN/CSA Std.C22.2 NO.60950-1 2nd Edition
User-added image
 
cETLus-Cert-4007497 ANSI/AAMI ES60601-1:2005 with AMD C1: 2009, AMD C2: 2010 and CAN/CSA-C22.2 No. 60601-1 2008 with COR 2:2011+ANSI/AAMI HA60601-1-11 Ed:1
User-added imageCHINA SJ/T 11364-2014, China RoHS Chart: http://en.globtek.com/globtek-rohs.php
Conforms to UL STD. 62368-1
Certified to CSA STD C22.2 NO.62368-1
Conforms to UL STD. 62368-1 Certified to CSA STD C22.2 NO.62368-1
User-added image  Declaration ДС № ЕАЭС N RU Д-US.КА01.В.10453_19 Custom Union of Russia, Belarus and Kazakhstan http://www.globtek.com/redirect/?loc=gost-certificate-eac-declaration
User-added image
FCC PART 15 CLASS B
User-added imageFuse 60335 T2.0A 250VAC Up to 36 volts
User-added imageIndoor Use Only - Mark is on the label or Molded in the case
 User-added image

Efficiency: complies to section 301 of Energy Independence and Security Act (EISA) complies with Energy Star tier 2 (North America), ECP tier 2 (China), MEPS tier 2 (Australia), Code of Conduct (Europe)
User-added imageNEMKO IEC/EN 60335-1 (Safety of Household and Similar Electrical Appliances, Part 1: General Requirements) up to 36 Volts
RoHSSpecifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf
User-added imageSemko S-Mark-Cert-EN60601-1 and EN60950-1
User-added imageUKCA Certification
User-added imageUkraine UKRSepro (Document: www.globtek.com/html/iso_certificates/GT_Ukraine.pdf)
WDaEAppliances can be used in warm damp equable climates
 User-added imageWEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm) Mark is on the label or Molded in the case
Popular Products (Stock items)
SKUVolts OutCurrent Out (Amps)ConnectorRFQSample Request
Vad är en open frame eller intern AC/DC strömförsörjning ? Open frame AC/DC strömförsörjningar är ochså känd som AC/DC omvandlare och är en typ intern strömförsörjning. Open frame strömförsörjningen är byggd in i produkten som önskas få ström och användas till att mvandla AC ström till DC spänning, Vad är ett IEC 62368-1 : 2018 power supply ? IEC 62368-1 : 2018 användas till säkerhet för elektriska och elektroniska apparater angående audio, video, information och kommunikationsteknologi, business- och kontormaskiner med en graderad spänning som ej överstiger 600V. Detta är en produktsäkerhetsstandard som klassifiserar energikällor, föreskrivna skydd mot dessa energikällor och ger information vid användning av- och krav till dessa skydd. Föreskrivna skyddskrav är menade för att minska sannolikhet för smärta, injurier och i fall av brand och egendom skador. Et IEC 62368 godkänd AC/DC power supply, är ett adapter som är testad och auktoriserad enligt bestämda standard. Ersätter ett IEC 62369-1 godkänd power supply ett IEC 60950-1 godkänd power supply ? 20 december 2020 är säkerhets- standardene IEC 60950-1 och IEC 60065 historia och ersättes av IEC 62368-1angående tekniska - och audio video produkter. I nord-Amerika vill ETL och UL 60950-1 godkanningar fortfarande existera och vill antagligen användas ytterligare. Hur fungerar ett AC/DC adapter ? AC/DC adapteren eller exsternt power supply är bestående av ett power supply som är skyddad i ett plasthus motsvarande en AC plugg, därmed namn som plug type, wall plug-in , plug in eller plugtop adapter. Adapteren monteras i vägguttaget och
Begär information
Call
Email
💬
Text
💁
Chat
?
Begär en offert
Kan vi hjälpa dig?
Kan vi hjälpa dig?